Technical Specifications
Page 1 of 1 © 2011 Multi Channel Systems MCS GmbH
USB-MEA32-STIM4-System
The USB-MEA32-STIM4-System is an amplifier with integrated
stimulus generator, filter amplification, and analog / digital board
converting analog signals in digital data streams in real time.
Technical Specifications
Operating temperature 10 °C to 50 °C
Storage temperature 0 °C to 50 °C
Relative humidity 10 % to 85 %, non-condensing
Dimensions (L x D x H) 235 x 135 x 55 mm
Weight 900 g
Amplifier
Gain 1000
Bandwidth 0.5 Hz - 3 kHz
Data resolution 16 bit
Number of recording channels 32
Input voltage range +/- 4096 μV
Cross talk (channel to channel) typical 0.01 %, max. 0.1 %
Data converter and USB interface
Sampling frequency up to 50 kHz
Maximal data rate 6.4 MByte per second
Data resolution 16 bit
USB Version USB 2.0 High Speed
Stimulus Generator
Output current +/- 256 μA
Resolution 13 bit
Maximum frequency 50 kHz
Time resolution 20 μs
PH01 perfusions cannula
Temperature senor type PT 100 with 4 wire connection
Accuracy +/- 0.1 °C
Heating element impedance 20
Interface and connectors
USB USB 2.0 High Speed cable (type A–mini B)
Power Supply barrel connector 0.7 x 2.35 mm
Ground common jack 4 mm
Power supply unit (MPU30)
Input voltage 90 – 264 VAC @ 47 – 63 Hz
Output voltage 11 – 13 V
Max. Power 30 W
Software
Operating system Microsoft Windows ® 7, XP or Vista
with NTFS
English and German version supported
MC_Rack program Version 3.6.6 and higher
Data acquisition and analysis software
MC_DataTool program Version 2.4.3 and higher
Data export ASCII file (*.txt) binary file (*.raw) format
MC_Stimulus II Version 3.0.0 and higher
Stimulation program
LTP-Director Version 1.0.0 and higher
Data acquisition, analysis and stimulation program
Warning: The USB-MEA32STIM4 device may only be used for the specified purpose. Damage of the device and even fatal injuries can result from improper use.