MEA Amplifier with Blanking Circuit for Upright Microscopes Manual

Seite 2

MEA Amplifier with Blanking Circuit for Upright Microscopes 10 3.2 Installing the Software System requirements Software: One of the following Microsof

Seite 3 - Table of Contents

First Use of the MEA Amplifier 11 3.3 MEA_Select Program MEA Amplifier with blanking circuit (MEA1060-BC) When starting the software program MEA_Sele

Seite 4

MEA Amplifier with Blanking Circuit for Upright Microscopes 12 A blanking signal (TTL pulse) transiently switch off the input stage of the amplifier d

Seite 5 - 1 Introduction

First Use of the MEA Amplifier 13 The command List Mode opens an additional window: List of Stimulation Patterns. In List Mode you can download a list

Seite 6

MEA Amplifier with Blanking Circuit for Upright Microscopes 14 3.4 Setting Up and Connecting the MEA Amplifier Warning: Spilled liquid can damage or

Seite 7 - 2.2 Guarantee and Liability

First Use of the MEA Amplifier 15 Please see also the separate handout "MEA Microelectrode Systems" for setup suggestions with detailed illu

Seite 8 - 2.3 Important Safety Advice

MEA Amplifier with Blanking Circuit for Upright Microscopes 16 Connecting the amplifier to the stimulus generator The following illustration shows the

Seite 9

17 4 First Tests and Tutorial 4.1 First Functional Tests Each MEA amplifier has been thoroughly tested at the factory site before delivery. However

Seite 10 - 3.2 Installing the Software

MEA Amplifier with Blanking Circuit for Upright Microscopes 18 The following screen shot shows a recording from a typical MEA amplifier with a test m

Seite 11 - 3.3 MEA_Select Program

First Tests and Tutorial 19 4.3 Computer Connection Testing the computer connection The MEA_Select program automatically detects and lists all serial

Seite 12

Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted without the express wri

Seite 13

MEA Amplifier with Blanking Circuit for Upright Microscopes 20 4.4.1 Operating multiple MEA Amplifiers with one STG It is possible to operate more th

Seite 14

First Tests and Tutorial 21 To start and stop only the channels associated with one amplifier, use the Start / Stop buttons for individual triggers. E

Seite 15

MEA Amplifier with Blanking Circuit for Upright Microscopes 22 1. Under Stimulation, select Ground. 2. Click any electrodes that you like to ground

Seite 16

First Tests and Tutorial 23 Stimulation without blanking Usually, you will always use the blanking when you stimulate. But it is possible to switch of

Seite 17 - 4 First Tests and Tutorial

MEA Amplifier with Blanking Circuit for Upright Microscopes 24 The stimuli are sent to the selected stimulating electrodes. You see stimulus artifacts

Seite 18

First Tests and Tutorial 25 You see that there are no stimulus artifacts visible on almost all non-stimulating electrodes, with some residual artifac

Seite 20

27 5 Operating the MEA Amplifier 5.1 Signal Amplification and Filters The MEA amplifier with blanking circuit is a 60-channel preamplifier with a br

Seite 21

MEA Amplifier with Blanking Circuit for Upright Microscopes 28 For information on the gain and filters of the MC_Card data acquisition system, please

Seite 22

Troubleshooting 29 DC offset correction The internal baseline restore routine of the preamplifier restores the offset to the ground level, that is, to

Seite 23

iii Table of Contents 1 Introduction 5 1.1 About this Manual 5 2 Important Information and Instructions 7 2.1 Operator's Obligations 7

Seite 24

MEA Amplifier with Blanking Circuit for Upright Microscopes 30 5.3 Mounting the MEA Probe and Grounding the Bath Warning: Do not use too much force.

Seite 25

Troubleshooting 31 1. Attach the provided silver wire or pellet to the amplifier's ground and place it into the bath, near the rim of the cultur

Seite 26

MEA Amplifier with Blanking Circuit for Upright Microscopes 32 1. Under Stimulation, select Ground. 2. Click any electrodes that you like to grou

Seite 27

Troubleshooting 33 5.5 Stimulation Recommended stimulus protocols for MEA electrodes MCS recommends to use only biphasic voltage pulses with the neg

Seite 28

MEA Amplifier with Blanking Circuit for Upright Microscopes 34 Important: You can only stimulate during an active TTL pulse, which is used for trigger

Seite 29 - 5.2 Temperature Control

Troubleshooting 35 If the List Mode is open, there is an additional button Trace List. If the Trace List button is active all other funct

Seite 30

MEA Amplifier with Blanking Circuit for Upright Microscopes 36 5.6 Stimulus Artifact Suppression (Blanking) A TTL pulse (blanking signal) that has to

Seite 31

Troubleshooting 37 Wait settings A delay (up to 5000 μs) between stopping stimulation (end of TTL pulse) and restarting the recording can be programme

Seite 32

MEA Amplifier with Blanking Circuit for Upright Microscopes 38 Single pulses versus repeated pulses Interestingly, the behavior of MEA electrodes when

Seite 33 - 5.5 Stimulation

Troubleshooting 39 Behavior of stimulation electrodes Due to the high electric charge that is applied by the stimulus and the general slow kinetics of

Seite 34

MEA Amplifier with Blanking Circuit for Upright Microscopes iv 6 Troubleshooting 44 6.1 About Troubleshooting 44 6.2 Error Messages 44 6.3 No

Seite 35

MEA Amplifier with Blanking Circuit for Upright Microscopes 40 Stimulus artifacts with switched off blanking on a typical non-stimulating 30 μm electr

Seite 36

Troubleshooting 41 60MEA200/10iR-Ti Voltage driven stimulation (1 V, electrode 33) on 10 μm TiN electrodes. The stimulus artifact suppression is slig

Seite 37

MEA Amplifier with Blanking Circuit for Upright Microscopes 42 Voltage driven stimulation (1 V, electrode 33) on 100 μm gold electrodes. You see only

Seite 38

Troubleshooting 43 The response of the cells to the stimulation is instantaneous in the cells near to the pulse. The cells recorded in the periphery

Seite 39

MEA Amplifier with Blanking Circuit for Upright Microscopes 44 6 Troubleshooting 6.1 About Troubleshooting The following hints are provided to solve s

Seite 40

Troubleshooting 45 MEA electrode is defective See also the MEA manual. Possible causes: ? The contact pads are contaminated.  Clean the contact pad

Seite 41

MEA Amplifier with Blanking Circuit for Upright Microscopes 46 6.5 Artifacts on All Channels You see strange artifacts on all channels. This behavior

Seite 42

Troubleshooting 47 6.6 Unpredictable Noise and Artifacts You use the amplifier for stimulation and you see unpredictable noise and / or artifacts on

Seite 43 - 5.8 Service and Maintenance

MEA Amplifier with Blanking Circuit for Upright Microscopes 48 Due to the higher electrode impedance, the maximum input voltage of the MEA1060-BC amp

Seite 44 - 6 Troubleshooting

Troubleshooting 49 6.8 Technical Support Please read the Troubleshooting part of the manual or help first. Most problems are caused by minor handling

Seite 45 - 6.4 Unsteady Baseline

5 1 Introduction 1.1 About this Manual This manual comprises all important information about the first installation of the hardware and software, a

Seite 47

51 7 Appendix 7.1 Technical Specifications 7.1.1 Pin and MEA Layout The layout of standard MEA electrodes follows the scheme of a standard grid: The

Seite 48

MEA Amplifier with Blanking Circuit for Upright Microscopes 52 Assignment pin to electrode 68-pin socket MEA electrode 68-pin socket MEA

Seite 49 - 6.8 Technical Support

Appendix 53 7.1.2 MEA1060-Up-BC Operating temperature 10 °C to 35 °C Storage temperature 0 °C to 50 °C Dimensions (W x D x H) 165 mm x 165 mm x

Seite 50

MEA Amplifier with Blanking Circuit for Upright Microscopes 54 7.1.3 FA60S-BC Operating temperature 10 °C to 50 °C Storage temperature 0 °C to 5

Seite 51 - 7 Appendix

Appendix 55 7.2 Contact Information Local retailer Please see the list of official MCS distributors on the MCS web site. User forum The Multi Channel

Seite 52

MEA Amplifier with Blanking Circuit for Upright Microscopes 56 USB-MEA-Systems Product Product Number Description USB-MEA-System for upright microsco

Seite 53 - 7.1.2 MEA1060-Up-BC

Appendix 57 USB-MEA-System with blanking circuit for inverted microscopes with advanced perfusion USB-MEA240-Inv-4-BC-System-E electrodes, 4 analog c

Seite 54 - 7.1.4 Test Model Probe

MEA Amplifier with Blanking Circuit for Upright Microscopes 58 Accessories Product Product Number Description Adapter for MEAs with 8 special electr

Seite 55 - 7.3 Ordering Information

Appendix 59 Spare Parts Product Product Number Description Heating element with 8 mm bore hole HE-Inv-8 Heating element with 12 mm bore hole HE-Inv-1

Seite 57

MEA Amplifier with Blanking Circuit for Upright Microscopes 60 MEA-Systems Product Product Number Description MEA recording system for inverted micr

Seite 58

61 8 Index

Seite 59

MEA Amplifier with Blanking Circuit for Upright Microscopes 62

Seite 60

63 3 3D-MEA 37 A Aliasing 25 B Band pass band...25 Bandwidth 25 C Change MEA 28 COM serial port ...

Seite 62

7 2 Important Information and Instructions 2.1 Operator's Obligations The operator is obliged to allow only persons to work on the device, who

Seite 63

MEA Amplifier with Blanking Circuit for Upright Microscopes 8 2.3 Important Safety Advice Warning: Make sure to read the following advice prior to i

Seite 64

First Use of the MEA Amplifier 9 3 First Use of the MEA Amplifier 3.1 Welcome to the MEA Amplifier Raw data from up to 60 electrodes of a microelec